The AS450 basic is protected by a device fuse .
The external fusing must be T5.0A (230V) F15A (110V).
• Turn the device off at the mains (H).
• Pull the mains plug from the low heat device socket (J).
• Unscrew the fuse insert (I) and remove the fuse (IS).
CAUTION
A fuse must be replaced by one of the same type and with the same fuse value.
• Place the new fuse in the fuse insert.
NOTE
Do not tighten the fuse insert too much when screwing the fuse in!
• Screw the fuse insert (I) back in.
NOTE
If the fuse triggers again immediately after being replaced, there is a fault in the
device which must be rectified.
There are two fuses: one in the live conductor and one in the neutral conductor of
the power supply.